Roomer S
Charger
The indicator on the lift flashes red when the
batteries need recharging. Plug the charger
cable into the hand control or put the hand
control into the charger cradle and charging
will start. When the lift is charging the indicator
glows yellow.
The lift can only be used with a
charger recommended by Human
Care, article number 50880.
• C harge the lift regularly, preferably every
night
• The charger must be disconnected when you
use the lift
• T he lift's function is blocked during charging.
Disconnect the charger to use the lift
Detaching the lift from the rail system
Lower the lift so that it stands solidly on a
surface. The lift will stop automatically when
it reaches the surface. It may be useful to use
Human Care's cart (Art.no. 50346). Feed more
belt out by pulling gently on the belt that is ac-
tive while at the same time pressing the Down
button on the hand control or control panel.
When the belt is slack it can be easily unhooked
from the distance strap by lifting the belt loop
upwards.
32
www.humancaregroup.com
• T he unit charges regardless of the position of
the main switch
• I f the emergency stop is activated the lift
cannot be charged. Check that the emergen-
cy stop is not activated when charging.
• The charger must only be connected to an
earthed socket.
Important! When the batteries
reach a critically low battery level,
it will no longer be possible to
perform more lifts (up or down).
However, there is still enough
battery power for emergency
lowering. See separate manual for
the charger, 99599.
Hold the active belt and turn the lever (10) to
the opposite position to release the pulled-out
belt. The belt will automatically be retracted.
Make sure the belt is kept taut and gets rolled
up in the lift straight, without folding.