Troubleshooting
Description
Nothing happens when you press the Up
button or the Down button on the hand
control or the control panel
Nothing happens when you press the Up
button or the Down button on the hand
control
Nothing happens when you press the Up
button or the Down button on the hand
control or the control panel.
Nothing happens when you press the Up
button or the Down button on the hand
control or the control panel.
The lift goes in the wrong direction when
you press the Up button or the Down
button on the hand control or the control
panel.
Nothing happens when you press the Up
button on the hand control or the control
panel.
Nothing happens when you press the
Down button on the hand control or the
control panel.
The belt is not fed out when the lift is on
a surface.
The indicator switches from flashing
green or yellow to flashing red.
34
Indicator
Action
Not lit
1) Check that the mains switch is on.
2) The batteries are fully discharged. Charge the lift
immediately
3) The emergency stop is pushed in. Deactivate by turning
the red button clockwise.
Flashing green
Check the hand control. If the hand control is defective it
needs to be replaced. Contact a service technician.
Flashing yellow
Flashing red
Constant red
1) Lift motor is overheated. Wait a while and try again.
2) Lift is overloaded. Lower using the emergency lowering
button
Constant yellow
Charging in progress. Interrupt charging.
Flashing green
The lever (10) has been turned in the wrong direction.
Constant green
Turn the lever towards the connected belt.
Flashing yellow
Flashing red
Flashing green
The lift is in its maximum upper position. Press the Down
button on the hand control or the control panel to lower
Flashing yellow
the lift.
Flashing red
Flashing green
The lift is in its maximum lower position. Press the Up
button on the hand control or the control panel to raise
Flashing yellow
the lift.
Flashing red
Flashing green
The lift belt has to be pulled upwards at the same time as
the Down button on the hand control or the control panel
Constant green
is held pressed in.
Flashing yellow
Flashing red
Flashing red
The batteries need charging but you can still lift the
patient a few times. The lift should be set to charge as
soon as possible.
ENG