Indicator
The indicator light (8) has the following
indications, which are only accessible when the
lift is switched on.
Flashing
The lift is on and ready for use.
Battery level is Full.
green
The lift is on and ready for use.
Flashing
yellow
Battery level is Normal. It might
be a good idea to charge up the
lift.
Flashing
The lift is on and ready for use.
Battery level is Low. The lift must
red
be charged up immediately.
Constant
The lift is in operation or loaded.
green
Charger
The indicator on the lift flashes red when the
batteries need recharging. Plug the charger
cable into the hand control or put the hand
control into the charger cradle and charging
will start. When the lift is charging the
indicator glows yellow.
The lift can only be used with a
o
charger recommended by Human
Care, article number 50880.
• Charge the lift regularly, preferably every
night
• The charger must be disconnected when
you use the lift
• The lift's function is blocked during
charging. Disconnect the charger to use the
lift
• The unit charges regardless of the position
www.humancaregroup.com
The lift safety catch has engaged.
Constant
red
(See troubleshooting table)
Constant
The lift is charging. The lift's
function is blocked.
yellow
The lift's service interval has
Flashing
green/
passed. Service and annual
inspection must be performed
yellow
immediately.
Flashing
The lift's service interval has
passed. Service and annual
red/yellow
inspection must be performed
immediately. Battery level is
Low. The lift must be charged up
immediately.
of the main switch
• If the emergency stop is activated the
lift cannot be charged. Check that the
emergency stop is not activated when
charging.
• The charger must only be connected to an
earthed socket.
Important! When the batteries
o
reach a critically low battery level,
it will no longer be possible to
perform more lifts (up or down). However,
there is still enough battery power for
emergency lowering. See separate
manual for the charger, 99599.