Secondary Control Panel; Indicator - human care Altair 55200H Manuel De L'utilisateur

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Secondary control panel

The lift is equipped with a secondary control
panel located on the lift which can be used
as an alternative to the hand control. It is
equipped with Up button (2) and Down button
(3) for raising and lowering, an OFF/ON button
(1) for activation/deactivation of the lift and
an emergency lowering button (4) for quick
lowering the lift.
Press emergency lowering button and down
button at least 3 seconds simultaneously to
activate the emergency lowering. Older lifts
(before july 2016) may not have double key
command and delay.

Indicator

The indicator light has the following indica-
tions, which are only accessible when the lift is
switched on.
Flashing
The lift is on and ready for use. Batte-
green
ry level is Full.
Flashing
The lift is on and ready for use. Batte-
yellow
ry level is Normal. It might be a good
idea to charge up the lift.
Flashing
The lift is on and ready for use. Battery
red
level is Low. The lift must be charged
up immediately.
Constant
The lift is in operation or loaded.
green
Constant
The lift safety catch has engaged. (See
red
troubleshooting table)
Constant
The lift is charging. The lift's function
yellow
is blocked.
Flashing
The lift's service interval has passed.
green /
Service and annual inspection must be
yellow
performed immediately.
Flashing
The lift's service interval has passed.
red /
Service and annual inspection must be
yellow
performed immediately. Battery level
is Low. The lift must be charged up
immediately.
www.humancaregroup.com