Secondary control panel
The lift is equipped with a secondary control
panel located on the lift itself. This control panel
can be used as an alternative to the hand con-
trol. The secondary control panel is equipped
with an Up button (2) and a Down button (3)
for raising and lowering the lift. The control
panel is also equipped with an OFF/ON button
(1) for activation/deactivation of the lift and
an emergency lowering button (4) for quickly
lowering the lift.
Press emergency lowering button and down
button at least 3 seconds simultaneously to
activate the emergency lowering. Older lifters
may not have double key command and delay.
When the lift is loaded with less than approx.
40 kg it can be run at two speeds. If the Up
or Down button is pressed for longer than 3
Indicator
The indicator light has the following
indications, which are only accessible when the
lift is switched on.
The lift is on and ready for use.
Flashing
green
Battery level is Full.
Flashing
The lift is on and ready for
use. Battery level is Normal. It
yellow
might be a good idea to charge
up the lift.
The lift is on and ready for
Flashing
red
use. Battery level is Low.
The lift must be charged up
immediately.
www.humancaregroup.com
seconds, the lift will run in the faster mode. If
the lift is loaded with more than approx. 40 kg
it will only run in the slower mode. The force
required to operate the push buttons is 4.4N.
1 2
Constant
The lift is in operation or
green
loaded.
The lift safety catch has
Constant red
engaged. (See troubleshooting
table).
The lift is charging.
Constant
yellow
Flashing
The lift's service interval has
passed. Service and annual
green/
yellow
inspection must be performed
immediately.
The lift's service interval has
Flashing
red/yellow
passed. Service and annual
inspection must be performed
immediately. Battery level is
Low. The lift must be charged
up immediately.
ENG
3 4
ENG