Storage; Preparation For Installation - Agilent Technologies TV 303 Notice De Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 37

Storage

In order to guarantee the maximum level of performance and
reliability of Agilent Turbomolecular pumps, the following guidelines
must be followed:
If for any reason the shelf life time is exceeded, the pump has to be returned to
CAUTION!
the factory. Please contact the local Agilent Vacuum Sales and Service
representative for information.

Preparation for Installation

The TV 303 is supplied in a special protective packing. If this shows
signs of damage which may have occurred during transport, contact
your local sales office.
When unpacking the system, be sure not to drop it and avoid any
kind of sudden impact or shock vibration to it.
Do not dispose of the packing materials in an unauthorized manner.
The material is 100% recyclable and complies with EEC 94/62 and
successive modification to protect the environment.
TV 303 User Manual / 87-901-019-01
when shipping, moving and storing pumps, the following
environmental specifications should not be exceeded:
temperature range: -20 °C to 70 °C
relative humidity range: 0 to 95 % (non condensing)
the turbomolecular pumps must be always soft-started when
received and operated for the first time by the customer
the shelf life of a turbomolecular pump is 10 months from the
shipping date.
(A)
Instructions for Use

Storage

199/268
16

Publicité

Table des Matières
loading

Table des Matières