6 INSTALLATION AND CONNECTION

Steps
What to do
1
Select the suction ring and the head
2
Assemble the head onto the motor
3
Carefully inspect the epikeratome
head after assembly
4
Check for separator oscillation
5
Inspect the suction ring
6
Set the stopper
7
Connect the suction ring to the
aspiration tubing

IMPORTANT WARNINGS

• The disposable head (#19390) is supplied sterile and
for single use only.
• Check that package is undamaged, unopened and
within the expiration date.
• Open the package containing the disposable head
(#19390).
• Screw the head clockwise onto the threaded motor
shaft.
• Check that the head is screwed on tightly to the motor.
There should be no rotation of the head.
• If the head is not screwed completely onto the motor,
the shaft will not drive the separator holder, and the
separator will not move. Poor or irregular resections
can result from incorrect assembly.
• Do not use the wrench (#19345C) or other tools
to assemble it. The wrench should be used for
disassembly only.
• The disposable head must be perfectly clean inside
and outside and free from debris, particles, oxidation
and deposits. There should be no scratches on the
plates. If any, replace the head.
• After assembly and using a microscope, carefully
inspect the separator under high magnification to
ensure that it is not damaged
• Start the motor to check proper blade oscillation.
• Do not use the Epi-K™ system if the oscillation is not
smooth, regular and uninterrupted.
• The suction ring (#19389/xx, 19399/-1) must be
carefully inspected prior to use; it must be perfectly
clean, free from debris, particles and deposits.
• Carefully check that the suction port is not obstructed.
• Refer to the indicative guidelines for the selection of
the ring and stop.
• The adjustable stop pin is pre-mounted on the suction
ring and used to determine the size of the hinge. The
hinge will be placed in a nasal position.
• To configure the stop, follow these steps:
o
unscrew the stop setting screw
o
lift it up then turn the stop to the appropriate
value
o
the value of the stop selected must line with the
etched ring size
o
screw the stop setting screw.
• The aspiration tubing (#19138) is supplied sterile and
for single use only.
• Check that package is undamaged, unopened and
within the expiration date.
• Inspect the aspiration tubing and in case of kinks or
obstructions, replace it.
#65043-H-02.2017