9.3 CLEANING- DISINFECTION-STORAGE OF THE HANDPIECE AND ELECTRICAL CABLE
Steps
What to do
1
Cleaning
• To clean the screw core of the motor and
metallic parts, use the disposable brush
(#19149) moistened with cleaning solutions.
• Dip the blade oscillation shaft of the motor in
a sterile distilled water bath and run the motor
back and forth 10 times.
• Remove the motor from the cup, unplug it from
the console, and wipe the oscillation shaft of
the motor with a lint-free cloth.
• Cable can be cleant with a lint-free cloth
moistened with cleaning solutions.
• Air-dry the shaft of the motor with clean, filtered
compressed air (medical compressed air).
2
Disinfecting & Drying
• Wipe each part of the motor with a lint-free cloth
moistened with disinfecting solutions.
• Use appropriate disinfecting solutions (spray
or wipe clean disinfectant) according to the
manufacturers' instructions.
• Dry carefully with disposable instrument cleaner
(lint-free) then with clean, filtered, compressed
air.
Storage
3
10
MORIA
#65043-H-02.2017
IMPORTANT WARNINGS
• The handpiece must be carefully cleaned
after each surgical procedure.
• Do not use abrasives or scrapers to clean
any elements of the system. Doing so may
result in reduced precision, burrs and/or
irregular tissue resection.
• Do not immerse the motor in the sterile
distilled water bath.
• To prevent damaging the connectors and
to ensure proper functioning of the motor,
never pull on the cables and never hold the
motor by its cable.
• It is imperative to verify that the devices are
perfectly clean and free of any organic or
other residues.
• Do not gas sterilize (ETO) the motor.
• Do not autoclave the motor.
• No current sterilisation process is
compatible with MORIA motors.
• Failure to completely dry the inside of the
motor can result in oxidation.
• Do not store products that are not
completely dry; doing so may create rust
and result in irregular corneal resection.
• When not in use, the Epi-K
be kept in its storage box (#22519512) in a
dry atmosphere.