Endosee
System - User Manual
®
•
Follow the instructions below to clean and disinfect
the Display Module. These instructions were
validated using CaviWipes
choose a wipe not listed here, you must validate it
before use.
•
Using a fresh, moist disinfectant wipe, thoroughly
wipe the front and back of the LCD screen of the
Display Module. Discard the wipe. Repeat with a
fresh wipe for a minimum of thirty (30) seconds.
Discard the wipe.
•
Using a fresh, moist wipe, thoroughly wipe the grip
section of the Display Module. Discard the wipe.
Repeat with a fresh wipe for a minimum of ten
(10) seconds. Discard the wipe.
•
Using a fresh, moist wipe, thoroughly wipe the seams
of the Display Module for one (1) minute.
•
Using a fresh, moist wipe, thoroughly wipe all
surfaces of the Display Module for a minimum of
ten (10) seconds. Discard the wipe.
•
Allow the wiped and wetted Display Module to stand
for a minimum of three (3) minutes.
•
At the end of the three (3) minutes, use a sterile,
lint-free wipe saturated with sterile 70% IPA to
thoroughly wipe the entire Handheld Monitor for a
minimum of thirty (30) seconds.
•
Allow the device to air dry. Once dry, visually inspect
the Display Module for any residual soil. Repeat
cleaning/disinfection instructions if visible soil is
present.
•
Place the cleaned/disinfected device in an
appropriate dry storage area.