The Endosee® System is not designed for use in
environments in which strong interference may
impact the performance of the equipment.
•
The emissions characteristics of this equipment
make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11
class B is normally required) this equipment might
not offer adequate protection to radio-frequency
communication services. The user might need to
take mitigation measures, such as relocating or
re-orienting the equipment
•
If this occurs, the site of use should be surveyed
to determine the source of this disruption, and the
following actions may be taken to eliminate
the source:
- Remove, reorient or relocate the
interfering equipment
- Increase the separation between the interfering
equipment and the Endosee
- Incrementally turn off equipment in the vicinity
to identify the interfering device
•
Do not attempt to transport or ship the
Endosee
System without using proper
®
packaging to protect the product.
•
This product is intended for indoor use only.