Warranty /
Procedure in case of repair
The warranty period for Bluephase 20i is 3 years from the
date of purchase (battery 1 year). Malfunctions resulting
from faulty material or manufacturing errors are repaired
free of charge during the warranty period. The warranty
does not provide the right to recover any material or
non-material damage other than the ones mentioned. The
apparatus must only be used for the intended purposes.
Any other uses are contraindicated. The manufacturer does
not accept any liability resulting from misuse and warranty
claims cannot be accepted in such cases. This is particularly
valid for:
• damage resulting from improper handling, especially
incorrectly stored batteries (see Technical data: Transpor-
tation and storage conditions).
• damage to components resulting from wear under
standard operating conditions (e.g. battery).
• damage resulting from external influences, e.g. blows,
drop to the floor.
• damage resulting from incorrect set-up or installation.
• damage resulting from connecting the unit to a power
supply, the voltage and frequency of which do not
comply with the ones stated on the rating plate.
• damage resulting from improper repairs or modifications
that have not been carried out by certified Service
Centres.
In case of a claim under warranty, the complete apparatus
(handpiece, charging base, power cord and power pack)
must be returned, carriage paid, to the dealer or directly to
Ivoclar Vivadent, together with the purchase document. Use
the original packaging with the corresponding cardboard
inserts for transportation.
Repair work may only be carried out by a certified Ivoclar
Vivadent Service Center. In case of a defect that cannot be
rectified, please contact your dealer or your local Service
Center (see addresses on the reverse side). A clear descrip-
tion of the defect or the conditions under which the defect
occurred will facilitate locating the problem. Please enclose
this description when returning the apparatus.

Product specifications

Delivery form
1 Charging base, 1 Power cord, 1 Power pack, 1 Handpiece 1 Battery (Li-Po),
1 10>8 mm light probe, 1 Anti-glare shield, 3 Anti-glare cones, Sleeves,
1 Set of Instructions for Use
Accessories
The following accessories are available for Bluephase 20i:
REF
627389
608538
551756
592496
627299
627300
627298
607922
608554
Technical data
Operating voltage charging base 5 VDC
Operating voltage handpiece
Power Pack
Operating conditions
Temperature
Relative humidity
Ambient pressure
Dimensions of the charging base L=205mm; W=150mm; H=85mm
Weight of the charging base
Charging time
Power supply of the handpiece
Max. battery time
Light source
Wavelength range
LED-Class 2
Light intensity
Operation
Dimensions of the handpiece
Weight of the handpiece
Transportation and storage conditions:
Temperature
Relative humidity
Ambient pressure
Bluephase 20i has to be stored in closed, roofed rooms.
Protect the device from severe jarring.
Battery
– Do not store at temperatures above 40 °C / 104 °F (or 60 °C / 140 °F for a
short period). Recommended storage temperature 15-30 °C (59-86 °F).
– Always store the battery charged and
– not for longer than 6 months.
Description
Light probe 10>8 mm, black (G2)
Pin-point light probe 6/2 mm, black (G2)
Anti-glare cones
Anti-glare shield
Bluephase 20i charging base (G2)
Bluephase 20i battery (G2)
Bluephase 20i handpiece (G2)
Bluephase Meter
Bluephase sleeves (G2)
3.7 VDC with battery
5 VDC with power pack
100-240 VAC / 50-60 Hz /
max. 0.4 A / Output 5 VDC
TYPE 15.2630
Manufacturer: Friwo
+10 °C to +35 °C (+50 to +95°F)
30 % to 75%
700 hPa to 1060 hPa
250 g
approx. 2 h (with the battery empty)
Li-Po battery
approx. 45 min.
(with a new, fully charged battery)
Polywave
LED
®
385–515 nm
LED LIGHT
DO NOT STARE INTO BEAM
CLASS 2 LED PRODUCT
Max. 5.11 mW / Wavelength 385 - 515 nm
2,000 - 2,200 mW/cm
2
5 min. on / 6 min. off (intermittently)
L=260mm; W=42mm; H=120mm
(incl. battery) 225 g
-20 °C to +60 °C (-4 to 140°F)
10% to 75%
500 hPa to 1060 hPa
19