What if ...

Symbol
Causes
Electronic defect in the handpiece
or battery
Apparatus is overheated or under-
cooled
Electronic defect in the battery
Battery empty
Charging contacts contaminated
No display
Battery in safety mode
during
(integrated protective circuit)
charging
Power pack not connected or
LED of the
defective
charging
base is not
illuminated
18

Error rectification

Remove and reinsert the battery. If the error persists, replace the battery
with the power pack (Click & Cure).
If the error persists, please contact your dealer or your local Service
Center.
Allow the apparatus to cool down (or assume room temperature if it is
undercooled) and try again after a certain time.
If the error persists, please contact your dealer or your local Service
Center.
Remove and reinsert the battery. If the error continues to be displayed,
place the apparatus into the charging base. If the error persists nonethe-
less, replace the battery with the power pack (Click & Cure).
Please contact your dealer or your local Service Center.
Place the handpiece into the charging base. The corresponding message
(„Charging") will be displayed. If the battery is not charged, the contacts
have to be cleaned.
If the error persists, please contact your dealer or your local Service
Center.
Check if the handpiece has been correctly placed into the charging base.
Clean battery contacts. Information: „Charging Battery" briefly appears
in the display before the display returns to the stand-by mode.
If cleaning the battery contacts does not resolve the problem, remove the
battery from the apparatus. Charge the battery separately from the hand-
piece in the charging base for approximately 10 minutes.
If the error persists, please contact your dealer or your local Service
Center.
Check the battery contacts and if the power pack is correctly positioned
in the charging base or if the power pack is connected to the power
supply by means of the power cord (display on the power pack lights up
in green if it works correctly).
If the error persists, please contact your dealer or your local Service
Center.