Guidance And Manufacturer's Declaration - Electromagnetic Immunity - Ivoclar Vivadent Bluephase 20i Mode D'emploi

Masquer les pouces Voir aussi pour Bluephase 20i:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 30
Appendix
Guidance and manufacturer's declaration – electromagnetic emission
The Bluephase 20i is intended for use in the electromagnetic environment specified below. The customer or the user of the Bluephase 20i should
assure that it is used in such an environment.
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations / flicker emissions
IEC 61000-3-3

Guidance and manufacturer's declaration - electromagnetic immunity

The Bluephase 20i is intended for use in the electromagnetic environment specified below. The customer or the user of the Bluephase 20i should
assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast
± 2 kV for power supply lines
transient/burst
± 1 kV for input/output lines
IEC 61000-4-4
Surge
± 1 kV line - line
IEC 61000-4-5
± 2 kV line - earth
Voltage dips, short
<5 % U
T
interruptions and
(>95 % dip in U
voltage variations
40 % U
on power supply
T
(60 % dip in U
input lines
IEC 61000-4-11
70 % U
T
(30 % dip in U
<5 % U
T
(>95 % dip in U
Power frequency
3 A/m
(50/60 Hz) magnetic
field IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
90
Compliance
Electromagnetic environment - guidance
The Bluephase 20i uses RF energy only for its internal function. Therefore, its RF emissions are very low and
Group 1
are not likely to cause any interference in nearby electronic equipment.
The Bluephase 20i is suitable for use in all establishments, including domestic establishments and those
Class B
directly connected to the public low-voltage power supply network that supplies buildings used for domes-
tic purposes.
P < 75W (pass without test)
Class A
P < 75W (pass without test)
Complies
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line - line
no prot. earth
<5 % U
T
) for 0.5 cycle
(>95 % dip in U
) for 0.5 cycle
T
T
40 % U
T
) for 5 cycles
(60 % dip in U
) for 5 cycles
T
T
70 % U
T
(30 % dip in U
) for 25 cycles
) for 25 cycles
T
T
<5 % U
T
(>95 % dip in U
) for 5 sec
) for 5 sec
T
T
30 A/m
Electromagnetic environment - guidance
Floors should be concrete or ceramic tile. If floors are covered with synthetic material,
the relative humidity should be at least 30%.
Mains power quality should be that of typical commercial or dental environment.
Mains power quality should be that of typical commercial or dental environment.
Mains power quality should be that of typical commercial or dental environment. If the
user of the Bluephase 20i requires continued operation during power mains interruptions,
it is recommended that the Bluephase 20i be powered from an uninterruptible power
supply or battery.
Power frequency magnetic fields should be at levels characteristic of a typical location in
a typical commercial or dental environment.

Publicité

Table des Matières
loading

Table des Matières