Vercise™ DBS Directional Lead Directions for Use
3. Push the end of the Lead into the port until it stops. Hold the Lead in place while sliding the locking
lever to the locked (
4. Secure the OR Cable to the microdrive or stereotactic frame.
5. Attach the OR Cable Extension to the OR Cable.
6. Ensure that the External Trial Stimulator is off.
7. Connect the OR Cable Extension to the External Trial Stimulator socket labeled "CD" (Figure 6).
WARNING: Always turn the External Trial Stimulator off before connecting or disconnecting the cable assemblies to
prevent unexpected stimulation.
Figure 6: Connection of the OR Cable to the External Trial Stimulator
8. Verify impedances are acceptable by using the CP or Remote Control to measure impedances.
9. Evaluate Lead placement:
a. Evaluate whether stimulation is appropriate to reduce symptoms.
b. Adjust the Lead location or stimulation parameters if necessary.
Note: The stylet should remain in place throughout Lead insertion and adjustments.
WARNING: High charge density can cause permanent tissue damage. The Clinician Programmer will limit
stimulation parameters to safe values.
WARNING: Increasing the number of Lead penetrations increases the probability of hemorrhage. The necessity
for an acute Lead revision should be minimized using techniques of target localization, such as microelectrode
recordings and imaging, to correctly place the Leads on the first attempt.
10. Turn off the External Trial Stimulator.
WARNING: A sudden increase in stimulation may occur if External Trial Stimulator is ON while disconnecting the
OR Cables.
11. Disconnect the OR Cable and OR Cable Extension from the proximal end of the Lead.
12. Verify that the Lead has not moved from the desired location.
Vercise™ DBS Directional Lead
90930938-02 REV A 4 of 247
) position (Figure 5).
Figure 5: Securing the Lead into the OR Cable Connector