 The update period of data is less than 5 seconds,which is changeable
according to different individual pulse rate.
 Please read the measured value when the waveform on screen is equably
and steady-going. This measured value is optimal value. And the waveform
at the moment is the standard one.
 The problem of overrating would emerge when the patient is suffering
from toxicosis which caused by carbon monoxide, the device is not
recommended to be used under this.

Clinical restrictions

a. As the measure is taken on the basis of arteriole pulse, substantial pulsating
blood flow of subject is required. For a subject with weak pulse due to
shock, low ambient/body temperature, major bleeding, or use of vascular
contracting drug, the SpO
the measurement will be more sensitive to interference.
b. For those with a substantial amount of staining dilution drug (such as
methylene blue, indigo green and acid indigo blue), or carbon monoxide
hemoglobin (COHb), or methionine (Me+Hb) or thiosalicylic hemoglobin,
and some with icterus problem, the SpO
may be inaccurate.
c. The drugs like dopamine,procaine,prilocaine,lidocaine and butacaine
may also be a major factor blamed for serious error of SpO
d. As the SpO
value serves as a reference value for judgement of anemic
2
anoxia and toxic anoxia, some patients with serious anemia may also report
good SpO
measurement.
2
Notice:
For the details of correlative clinic restriction and contraindications, please refer
to the related medical literatures.

5.2 Menu Operation

Under any monitor condition, press
The diagram is showed as Fig. 5-3.
Fig. 5-3 Main menu
ENGLISH
waveform (PLETH) will decrease. In this case,
2
OK
determination by this monitor
2
to enter main menu interface.
measure.
2
58