Physical Characteristics - Gima CMS-VESD Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Chapter 3
Technical Specifications
Product name: Multi-functional Visual Stethoscope.
Safety: Complies with IEC 60601-1.
Classification:
EMC: Group I Class B.
Anti-electroshock Type: Internally powered equipment.
Anti-electroshock Degree: type B applied part.
(The oximeter probe is type BF applied part )
Harmful Liquid Proof Degree: IPX0.
Degree of Safety in Presence of Flammable Gases: Equipment is not suitable for
use in presence of flammable gases.
Mode: Continuous work.

Physical Characteristics

Size: 110 (length) × 60 (width) × 14 (height) mm
Weight: 100g(including battery)
Environment
Transport and storage
Temperature: -10°C~+50°C
Relative Humidity: 10%~95%
Atmospheric pressure: 50kPa~106kPa
Working environment
Temperature: 5°C~40°C
Relative Humidity: 20%~80%
Atmospheric pressure: 70kPa~106 kPa
Battery type: charging lithium battery
Working-voltage: DC3.6V ~ DC4.2V
Power: ≤120mA
LCD display: 2.4" TFT, 320x240 pixels
Continuous running time: 7 hours
Main Parameters:
1) Measurement of auscultation volume
heart sound range: 20Hz~230Hz
lung sound range:100Hz~800Hz
heart&lung combination sound range: 20Hz~800Hz
2) Measurement of Heart Rate
range: 30bpm~300bpm;
resolution: 1bpm;
accuracy: ±2 bpm or ±2% (select larger).
ENGLISH
52

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

32597

Table des Matières