3) Measurement of SpO
range: 0%~100%;
resolution: 1%;
accuracy: When the SpO
of absolute error is ±2%. below 70% unspecified.
4) Measurement of pulse rate
range: 30bpm~250bpm;
resolution: 1bpm;
accuracy: ±2 bpm or ±2% (select larger).
5) Measurement Performance in Weak Filling Condition:
SpO
and pulse rate can be shown correctly when pulse-filling ratio is 0.4%.
2
SpO
error is ±4%, pulse rate error is ±2 bpm or ±2% (select larger).
2
6) Resistance to surrounding light:
The deviation between the value measured in the condition of man-made light
and indoor natural light and that of darkroom is less than ±1%.