7 Troubleshooting

7.1 Troubleshooting
If your concentrator fails to operate properly, refer to the following chart for possible causes and solutions. If necessary,
contact your provider.
Oxygen Concentrator Does Not Turn On or Does Not Stay On
Symptom
Probable Cause
Concentrator does
On/Off button was not held down long enough.
not operate when
On/Off button is
pressed.
Battery is discharged (or improperly connected to
OR
the concentrator).
Concentrator
begins to operate
when On/Off
button is pressed,
but all LCD and
Battery is improperly connected to the concentrator. 1. Remove the battery and any external power
front panel lights
go dark.
Battery has performed a self-protect shutdown
when there was no external power.
48
Solution(s)
Try to power up the concentrator again while
continuing to press the On/Off button until the
front panel lights and LCD begin to illuminate. This
takes typically one full second.
1. Remove battery and check battery gauge.
2. If gauge shows battery is charged, then reinstall
battery and retry.
3. If gauge shows battery is depleted, then either
install a charged battery or connect external
power and retry.
connection.
2. Reinstall the battery, making sure it is
completely inserted and retry.
3. Connect to either AC or DC external power
and retry.
1. Wait one minute and retry.
2. Change to an alternate battery or connect
concentrator to either AC or DC external power
and retry.