WARNING!
Risk of Injury or Death
To reduce the risk of injury or death from illness:
– Replace the nasal cannula on a regular basis.
Check with your equipment provider or
physician to determine how often the cannula
should be replaced.
– DO NOT share cannulas between patients.
Cannulas are for single patient use only.
CAUTION!
Risk of Damage
– Use of sterilization or cleaning solutions may
leave this device nonfunctional.
CAUTION!
Risk of Minor Injury or Discomfort
To ensure proper breath detection and oxygen
delivery:
– Ensure the cannula prongs are positioned
properly in your nose. This is critical to the
effectiveness of the oxygen therapy.
– DO NOT use tubing/cannula length exceeding
25 ft (7,6 m).
– Use crush-proof oxygen tubing.
– Use ONLY single lumen cannulas with the
flow characteristics stated in Specifications in
Technical Data.
– Avoid use of concentrator in windy conditions.
– Check for gas flow at the outlet of the cannula.
1187376-C
1. Remove the cannula from its packaging.
2. Connect the cannula to the concentrator oxygen outlet
port A. Ensure the connection is secure.
3. Place the cannula B over your ears and position the
prongs in your nose as instructed by your health care
provider or cannula manufacturer.
Gas flow at the outlet of the cannula can be checked
while the concentrator is warming up. Wave your
hand in front of the nasal prongs. You should be
able to hear and feel the pulsed flow of gas. If
you do not feel the gas pulse, check the cannula
connection for leaks.
Setup
33