Invacare® Platinum® Mobile
Oxygen Purity:
Conserver Trigger
Sensitivity:
Conserver Breath Rate
Capacity:
Maximum Outlet
Pressure:
Cannula Requirements:
Battery Specifications
(each battery):
Battery Shelf Life:
66
87% to 95.6%, at all flow settings and over the rated ranges for ambient temperature, humidity
and atmospheric pressure.
After initial warm-up period (typically less than 5 minutes)
< 0.18 cmH
O pressure drop (For all cannula lengths)
2
Factory set—no adjustment, pressure activated
Only patient respiratory efforts that achieve the trigger pressure will result in the delivery of
an oxygen bolus.
15–40 BPM (breaths per minute) without reduction of bolus minute volume
28.5 psig (197 kPa)
Length: 4–25 ft (1.2–7.6 m) including all oxygen tubing
Tubing: crush-proof, single lumen
Adult, standard flow (rated for up to 6 L/min continuous flow) for lengths up to 7 ft)
Adult, high flow (rated for up to 15 L/min continuous flow) for lengths greater than 7 ft to 25 ft)
Example of Possible Cannula Model: Westmed Inc. Part Number 0194 (4 ft length)
Rechargeable lithium-ion, 14.4 V, 5800 mAh, 83.5 Wh, 500 full charge/discharge cycle life
12 months from date of manufacture
1187376-C