Invacare® Platinum® Mobile
No External Power Symbol Illuminated on Control Panel
Symptom
With AC power
Power source is faulty, or there is a loose
plugged into
connection.
concentrator,
external power
icon is not
illuminated on the
control panel.
With DC power
Power source is faulty or there is a loose
plugged into
connection.
concentrator,
The DC power cable fuse has blown.
external power
icon is not
illuminated on the
control panel.
7.2 Alarm Conditions
When any alarm condition occurs, the yellow alarm indicator light on the control panel illuminates and text describing
the alarm condition is shown on the display screen. Refer to the solution section of the alarm tables for potential
corrective actions. If necessary, contact your provider.
In all cases, pressing and holding the On/Off button for one second will turn off and reset the concentrator.
If the cause of the alarm condition is not corrected, it will reappear when the concentrator is turned back on.
All alarms are classified as Low Priority technical alarm conditions.
When multiple alarm conditions exist, the text associated with the highest ranked alarm will be shown on the display
screen.
50
Probable Cause
Solution
1. If the green indicator light on the AC adapter
is lit, check power connections from the AC
adapter to concentrator.
2. If the green indicator light on the AC adapter
is not lit, check power connections from outlet
to AC adapter or try another outlet.
1. Check power connections to concentrator.
2. Try another power outlet.
1. Switch to AC power and contact your provider
to service the DC power cable.
1187376-C