LCD DISPLAY TEXT:
Shut Down
System Fault
INDICATORS:
Two audible beeps every 16 seconds
YELLOW alarm indicator illuminated
NOTES:
To clear the alarm, turn the concentrator off. Restart concentrator.
Call your provider and report the problem if alarm continues.
LCD DISPLAY TEXT:
Not applicable. All power lost, display
is blank.
INDICATORS:
Audible beeps every second for
approximately 30 to 60 seconds
No other control panel indicators
illuminated
NOTES:
This alarm occurs even if the concentrator is off when all power is removed.
To clear the alarm, complete solutions 1, 2, or 3 before restarting.
1187376-C
DESCRIPTION:
Abnormal system condition has been
detected.
May be caused by malfunction of
compressor, fan, pressure or software.
Concentrator function has shut down.
DESCRIPTION:
All power sources have been
lost/removed.
Concentrator function has shut down.
SOLUTIONS:
1. Turn the concentrator Off, then On
again to retry.
2. Change to another source of oxygen
if alarm continues
SOLUTIONS:
1. Ensure the battery is completely
inserted.
2. Check external power connections
to the concentrator and to the
power outlet.
3. Try another power outlet.
4. Change to another source of oxygen
if alarm continues.
Troubleshooting