LCD DISPLAY TEXT:
Warming Up
INDICATORS:
No audible beep
YELLOW alarm indicator illuminated
NOTES:
Depending on a variety of factors, it can take up to 15 minutes to reach the minimum specified oxygen purity.
The alarm will turn off when the minimum oxygen output purity is reached.
If the minimum oxygen output purity has not been reached within 15 minutes from initial start up, the alarm will become a
Low Purity alarm condition.
58
DESCRIPTION:
The concentrator is still warming up
after two minutes from initial start
up. The oxygen output purity has not
reached the minimum value listed in
Specifications in Technical Data.
SOLUTIONS:
1. Verify concentrator is in
recommended environmental
temperature. Refer to Specifications
in Technical Data.
2. Clean intake filter screen and ensure
both intake and exhaust are not
blocked.
3. Allow concentrator to continue
warming up for at least 15 minutes.
4. Turn concentrator off, then on again
to retry.
1187376-C