Always ensure the concentrator battery(s) are recharged
as soon as possible after they become fully discharged.
The battery(s) may be permanently degraded if left fully
discharged for an extended length of time.
Check the status of your concentrator battery(s) once
a month if you are not using your concentrator on
a daily basis. Battery(s) should be maintained at two
illuminated bars worth of charge if not using the
concentrator on a daily basis.
Heat is the worst enemy of a battery. Allow plenty of
air to circulate around the concentrator so that the
battery(s) is kept as cool as possible when in use and
when charging.
What Not to Do
DO NOT use or leave the concentrator or battery(s)
in excessive heat or cold.
DO NOT store or leave the concentrator or battery(s) in
car trunks, etc. for extended periods of time.
DO NOT store battery(s) fully charged (four bars
illuminated on the unit's battery gauge) if you are going
to store your concentrator for any time greater than
a day. Recharge or discharge the battery to two bars
(50% charge) only. Storing a battery with a full charge
may degrade its useful life.
DO NOT leave your battery(s) installed into the
concentrator when the concentrator is not going to
be used for more than a day. The battery(s) will lose
charge while plugged into the concentrator even with
the concentrator turned off.
1187376-C
Usage
39