Invacare® Platinum® Mobile
LCD DISPLAY TEXT:
NOTES:
A sufficient reduction of the user's breath rate will automatically turn off the alarm.
The first occurrence of this alarm condition will cancel the audio off (mute) feature. Audio off (mute) can be reactivated. If
reactivated, the alarm audio off (mute) feature will remain activated even if this alarm condition occurs again.
LCD DISPLAY TEXT:
Low Battery
Charge Concentrator
INDICATORS:
Single audible beep no repeat
YELLOW alarm indicator illuminated
Battery gauge:
Pulse Flow Settings P1–P3: only the
lowest segment is illuminated
Pulse Flow Setting P4–P5: only
the lowest two segments are
illuminated
NOTES:
The alarm turns off when external power is connected or when a charged battery is installed.
Alarm will turn off when a breath is detected.
52
DESCRIPTION:
DESCRIPTION:
Remaining battery run time of
approximately 30 minutes. Battery
requires charging.
SOLUTIONS:
SOLUTIONS:
1. Connect oxygen concentrator to
either the AC or DC power or insert
an alternate charged battery into
the unit, OR
2. Insert an alternate charged battery
into the unit. Remove the depleted
battery and charge it with the
external battery charger for future
use.