Contra-Indications - arjo Enterprise 9600X Notice D'utilisation

Masquer les pouces Voir aussi pour Enterprise 9600X:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Contra-
The bed is not suitable for use in the following situations:
indications
The maximum recommended patient weight is 185kg (407 lb).
185 kg/
407 lb
The safe working load (SWL) of the bed is 250kg (550 lb).
250 kg/
550 lb
The safe working load is calculated as follows (in accordance
with IEC 60601-2-52):
Maximum patient weight.............................. 185kg (407 lb)
Mattress........................................................... 20kg (44 lb)
Accessories (including attached loads) ........... 45kg (99 lb)
TOTAL......................................................... 250kg (550 lb)
WARNING
If the combined weight of the mattress and accessories exceeds
65 kg, the maximum patient weight must be reduced accordingly.
The recommended patient size is; weight equal or above
40kg (88 lb), height between 146cm (57 ins.) and 190cm (75 ins.)
and BMI equal or above 17.
At the discretion of the carer, patients taller than 190cm may be
146
cm
BMI 17
40
kg
accommodated by extending the bed - refer to "Bed length
adjustment" on page 27. Ensure that the patient's height does not
exceed the In-bed length shown on page 68.
• A domestic area, i.e. home healthcare.
• Outpatient care.
• By adult patients under 40kg (88 lb) in weight.
• By children under 12 years old.
11

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Enterprise 9700xEnterprise 9800x