8. Technical Data
General
Safe working load
Maximum patient weight
Product weight (approx.)
Audible noise
Operating conditions
Temperature
Relative humidity
Atmospheric pressure
Electrical data
Power input
Duty cycle
Safety standards USA/Canada
Electric shock protection
EMC
Potential equalisation terminal
Liquid ingress protection
Backup battery
Patient weighing system
Minimum verification interval
(scale division) e
Minimum capacity
Maximum capacity
250 kg (550 lb)
185 kg (407 lb)
180 kg (396 lb)
50 dB approx.
10°C to 40°C (50° F to 104 °F)
20% to 90% at 30°C (86 °F), non-
condensing
700hPa to 1060hPa
3A max. at 230V a.c. 50/60Hz
3A max. at 230V a.c. 60Hz (KSA)
5.8A max. at 120V a.c. 60Hz
10% (2 min. on, 18 min. off)
EN/IEC 60601-1:2005 AMD1:2012
ANSI/AAMI ES60601-1 (2005)
AMD 1 (2012)
CAN/CSA-C22.2 No. 60601-1:14
IEC60601-2-52:2015
Class I
Type B
Complies with IEC 60601-1-2:2014
Complies with EN/IEC 60601-1:2005
AMD1:2012
IPX4
2 x 12V series connected, sealed,
rechargeable lead/acid gel, 1.3Ah
500g (1 lb)
10kg (20 lb)
250kg (550 lb)