The bed is not suitable for use in the following situations:
indications
The maximum recommended patient weight is 185kg.
The safe working load (SWL) of the bed is 250kg.
The safe working load is calculated as follows (in accordance with
IEC 60601-2-52):
Maximum patient weight ...................................................185kg
Mattress ...........................................................................20kg
Accessories (including attached loads) ..............................45kg
TOTAL
WARNING
If the combined weight of the mattress and accessories exceeds 65kg, the maximum
patient weight must be reduced accordingly.
The recommended patient size is: weight equal or above 40kg, height
between146cm and 190cm, and BMI equal or above 17.
At the discretion of the carer, patients taller than190cm may be
accomodated by extending the bed - refer to "Bed length adjustment" in
146
cm
BMI 17
40
kg
Chapter 4. Ensure that the patient's height does not exceed the "In-bed
length" shown in Chapter 7.
• A domestic area, i.e. home healthcare.
• Outpatient care.
• By patients under 40kg in weight.
• By children under 12 years old.
.........................................................................250kg