WARNING
Connection to the nurse call and RS232 connector must be made with
the correct cables. Use of the wrong cables may cause malfunction of
the bed or devices connected to the nurse call or RS232 connectors.
Verify correct operation of the nurse call and RS232 system before
placing a patient on the bed.
Do not allow the nurse call or RS232 cables to trail on the floor where
they may cause a trip hazard.
Make sure the nurse call and RS232 cables are disconnected before
moving the bed.
Do not allow the cables to touch the bed frame as this may affect the
accuracy of the weighing system.
WARNING
Only connect devices designed to be used with the bed. Connecting
devises not designed to be used with the bed may result in damage to
the equipment or to the bed.
NOTE
Only use cables with an in-line breakaway connector (below).
Failure to do so could result in damage to the bed or to other hospital equipment.
Fig. 6 - Example of an in-line breakaway connector
16
Breakaway
connection
Connect to
bed
Connect to
Nurse call system