arjo Enterprise 9600X Notice D'utilisation page 63

Masquer les pouces Voir aussi pour Enterprise 9600X:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Symptom
Error code E410 and
flashing control panel
lights
Alarm sounds when
patient moves on the bed
Errors in indicated patient
weight
Nurse call function not
working
Backrest does not lower
using the manual CPR
release handle
Error code E001
Error code E002
Error code E003
Error code E102
Possible Cause
Flat battery
Movement detection
threshold set too low
Mattress platform
obstructed
Nurse call cable not
connected or wrong type
Insufficient patient weight
applied to the backrest
Zero on battery power
Auto Compensation
weight decrease more
than 100kg (220 lb.)
Auto Compensation
weight increase more than
100kg (220 lb.)
A stable weight reading
was not obtained within 10
seconds
Action
If the bed has just been
delivered or has been
left in storage for a long
period of time without
charge, plug the bed
into mains power and
leave to charge as per
instructions.
If the problem persists
call an Arjo approved
service engineer
Increase VariZone patient
movement detection
threshold setting
Check the mattress
platform is not touching
furniture, curtains, cables,
etc.
Check that weight has not
been added to the bed
without using the auto-
compensation function.
Check cable is of correct
type and properly
connected
Push the backrest down to
initiate lowering
Connect bed to mains
supply and re-initialise
the weighing system -
see page 45
Replace weight removed
from bed
Remove added weight
from bed
Refer to the points listed
in the section
"Accuracy" on page 44
63

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Enterprise 9700xEnterprise 9800x

Table des Matières