End of life disposal
• Equipment that has electrical and electronic components should be
disassembled and recycled per Waste of Electrical and Electronic Equipment
(WEEE) or in accordance with local or national regulation.
• All batteries in the product must be recycled separately. Batteries are to be
disposed in accordance with national or local regulations.
• Components that are primarily made up of different kinds of metal (containing
more than 90% metal by weight) for example bed frame, should be recycled as
metals.
Transport and storage
Handle with care. Do not drop. Avoid shock or violent impact.
This equipment should be stored in a clean, dry and well-ventilated area which
meets the following conditions:
Temperature
Relative humidity
Atmospheric pressure
Caution
If the bed is stored for a long time, it should be connected to the
electricity supply for 24 hours every three months to recharge the
backup battery, otherwise it may become unserviceable.
Symbols
250 kg/
550 lb
185 kg/
407 lb
-10°C to 50°C (14°F to 122°F)
20% to 90% at 30°C (86 °F), non-condensing
700hPa to 1060hPa
Safe working load
Maximum patient weight
Alternating current (a.c.)
Caution
Refer to instructions for use
Type B applied part
Applied parts are considered to be: Upper frame
section, Bed controls, Split side rails, Head and Foot
Boards
69