34) Appuyer sur le bouton EXIT (QUITTER) pour fermer la FENÊTRE DE POST-RECONSTRUCTION.
6.1. 2D SINGLE SHOT
In the Acquisition windows pressing "2D single shot" button you can carry out one or more scout view at the
following angles:
0° 45° 90° 135° 180° 225° 270° 315°
1) After pressing the button
, the following window appears:
2) You can select one or more scouts by clicking on the images; the selecting ones appear with green edges, then
you have to click on the "Prepare" button.
3) Position the patient and arm the WhiteFox; when the bar status becomes green, press the X-ray exposure
button on the WhiteFox remote control to acquire a scout; please see the previous chapter about the scout view
procedure.