PROCEDURE 3. RECORD THE INTUBATION
Please read the
1.
Start the recording by ensuring the following conditions are met:
A blade or video baton is connected to the monitor.
A USB flash drive is connected to the micro‑USB port on the monitor.
The monitor is powered on.
The record feature is turned on in the user settings.
Once these conditions are met, the recording starts automatically.
2.
Stop the recording by pressing and holding the power button until the monitor
has fully powered off. The recording also stops when the video laryngoscope is
disconnected, the media capacity on the USB flash drive drops too low, or the
monitor's remaining battery charge provides less than 1 minute of power.
3.
If you would like to review the recording, connect the USB flash drive to a computer,
and then view the .avi file. Files are automatically named with the system date and
time.
PROCEDURE 4. DISCONNECT THE BATON (VIDEO BATONS ONLY)
The GVL Stat is a sterile, single‑use device. After each use, it is a biohazard, and it should
be removed from the video baton and disposed of in a manner consistent with local
protocols.
1.
Hold the Stat in one hand.
2.
To reduce the force required to remove the video baton from the Stat, use your
thumb and finger to gently press the collar of the Stat.
3.
With the other hand, grasp the handle of the video baton and pull firmly.
Warnings & Cautions
section before performing the following task.