ELITE SII
Loading of the
programs from the
memory.
Battery charge.
In compliance with:
EN 60601-1: 2006 + A1: 2013
EN 60601-1-2: 2015
EN 60601-2-10: 2015
EN 60601-1-11: 2015
Warning: radiofrequency communication devices (including accessories like
antennas or antenna cables) must be used at least 3 meters away from every part
(including cables and accessories) of the device. Otherwise performance can be
affected.
REV 5 DEL 16.10.2023
GENESY SII
reconnection of the
electrode.
Setting of a
Dangerous
current that is
stimulation.
too high
Error in the
Execution of an
data from the
incorrect program.
memory.
Battery
Damaging of the
overheating.
device, display
information no
longer readable,
explosion, fire.
DUO TENS
current is below a certain
threshold, the machine must
rest the current of the
channel.
The device must derate the
voltage boost stage to
prevent emitting a current
beyond the maximum value.
The machine that does not
detect the electrode
contact must reset the
channel current and must
indicate this reset in the
indication on the display.
The machine must check
the correctness of the data
of the programs. In case an
error is detected, the device
must restart.
The device must monitor the
temperature of the battery,
if a certain threshold is
exceeded, the battery
charging must be
interrupted.