ELITE SII
PED1:
It is possible to set the values of frequency and duration of the pulse.
From 1 to 150 Hz in steps of 1 Hz.
From 50 to 250 µs in steps of 10 µs.
Default value: 100 Hz / 200 µs.
PED2:
It is possible to regulate the pulse duration with the frequency set on 100
Hz. From 50 to 250 µs in steps of 10 µs.
Default value: 200 µs.
PED 3:
It is possible to set the frequency values.
From 1 to 150 Hz in steps of 1 Hz.
Default values: 100 Hz
It is possible to modulate the pulse duration from 100 µs to 250 µs.
PED 4:
It is possible to set the values of frequency and pulse duration.
From 1 to 150 Hz in steps of 1
Hz.
From 50 to 320 µs in steps of 10
µs.
The following values are not editable: tup = 1 s; thold = 5 s; tdown = 1 s; toff = 5 s
Default values: 30 Hz / 250 µs.
t
t
UP
HOL
PED 5:
It is possible to set the frequency values.
From 1 to 150 Hz in steps of 1 Hz.
Default values: 30 Hz
It is possible to modulate the pulse duration from 100 µs to 320 µs.
PED 6
:
It is possible to set the values of t
REV 5 DEL 16.10.2023
GENESY SII
t
t
DOWN
OFF
HOLD
DUO TENS
e t
.
OFF
61