ELITE SII
are subjective and highly variable. It is recommended to avoid treatments in the
abdominal zone during menstruation cycle and immediately before or after it.
Is it possible to use the electrostimulator during lactation?
Until now, no collateral effects regarding lactation have been observed.
Nonetheless, during lactation, it is recommended not to stimulate the thoracic
region.
Are dermatological diseases (e.g. psoriasis, urticaria) contraindications for
electrostimulation?
Yes, do not treat areas affected by dermatological diseases.
When are the first results visible?
Aesthetic results of electrostimulation are always subjective. For Toning program,
with a regular program of 3-4 sessions per week, a noticeable result may be
observed after 15 days. For Lipolysis and Drainage programs, 40 days of treatment
are necessary. Results are obtained more quickly if treatments are combined with
good physical activity and a correct life style.
How many sessions can I weekly perform?
For physical training, consult the program of weekly training in the Globus Personal
Trainer. For fitness and aesthetics programs, the number of sessions depends on the
type of treatment: 3-4 sessions per week on alternate days are suggested for toning;
daily treatments are permitted for Lipolysis and Drainage programs.
EMC accompanying documents
Essential performance
PERFORMANCE
Electrostimulation.
REV 5 DEL 16.10.2023
GENESY SII
CONDITION
RISK
External
Display information
disturbance
no longer readable.
(Burst).
Lack of
Interruption of the
internal power
treatment.
supply.
Lack of
Interruption of the
external
treatment.
power supply.
Detachment
Unpleasant
of an
stimulation or
electrode.
painful electric
shock in case of
DUO TENS
ACCEPTED EVENT
The machine must stop the
stimulation.
The machine must maintain
the stimulation and accept
the commands.
The machine must signal the
battery exhaustion and the
interruption of the treatment.
The device, if equipped with
a battery, must continue the
treatment signaling that
operation is carry out in
battery mode.
The device must constantly
monitor the current on each
active channel set over 9
mA. In case the detected
82