Accessories and spare parts are available via the corresponding
service address. Be sure to specify the corresponding order code
when ordering.
Designation
Universal cuff (22‒42cm)
Error
message
Er1
Er2
Er3
Er4
Er5
460437_2401_V3.0_silv_Blutdruckmessgeraet_Content_OS.indd 65
460437_2401_V3.0_silv_Blutdruckmessgeraet_Content_OS.indd 65
Accessories and space parts
Troubleshooting
Problem
Pulse could not be
measured correctly.
The measured
blood pressure is
outside of the
measuring range.
A pneumatic system
error has occurred.
An error occurred
during the measure-
ment process.
The inflation pres-
sure is more than
300 mmHg.
Article number or order
number
164.492
Solution
Please repeat the measurement after a
one-minute break. Please refrain from talk-
ing or moving during the measurement
process.
Repeat the measurement. Ensure that the
cuff tubing has been connected correctly;
refrain from moving and talking.
Please repeat the measurement after a
one-minute break. Please refrain from talk-
ing or moving during the measurement
process.
Please complete a new measurement to
check whether the cuff can be properly
inflated. Make sure that neither your arm
nor heavy objects are placed on the tub-
ing and that the tube is not kinked.
GB/IE
65
29.05.24 10:51
29.05.24 10:51