Troubleshooting
5
Troubleshooting
5.1
Warnings
Fault
Cause
The residual run-
time of the
A segment of the
rechargeable bat-
rechargeable bat-
tery is low
tery status indica-
(< 30 minutes).
tor is flashing
yellow.
An acoustic signal
sounds (a low tone
2x/minute).
The particle filter
residual capacity is
A segment of the
low (< 20 %).
particle filter resid-
ual capacity indica-
tor is flashing
yellow.
An acoustic signal
sounds (a low tone
2x/minute).
Malfunction during
switch-on (e.g.
A segment of the
caused by missing
flow rate indicator
hose or filter).
is flashing yellow.
An acoustic signal
sounds (recurring,
high tone).
5.2
Alarms
5.2.1 Blower unit
Fault
Cause
The residual run-
time of the
A segment of the
rechargeable bat-
rechargeable bat-
tery is almost
tery status indica-
exhausted
tor is flashing red.
(< 10 minutes)
An acoustic signal
sounds (recurring,
high tone).
The particle filter
residual capacity is
A segment of the
almost exhausted
particle filter resid-
(< 10 %)
ual capacity indica-
tor is flashing red.
An acoustic signal
sounds (recurring,
high tone).
Faulty breathing air
supply during oper-
A segment of the
ation (e.g. caused
flow rate indicator
by a missing hose,
is flashing red.
missing filter, or
An acoustic signal
kinked hose).
sounds (recurring,
high tone).
22
Fault
One segment at a
Remedy
time is flashing red.
Recharge the bat-
An acoustic signal
tery soon or
sounds (recurring,
replace with fully
high tone).
charged battery.
5.2.2 Standard battery charger
Fault
Status LED is
Change particle or
flashing red.
combination filter
soon.
6
Re-check the
device function
and prepare for
6.1
use.
Work to do
Clean and disinfect the device
Carry out a visual inspection (see "Visual
Remedy
inspection", page 23)
Recharge the bat-
Carry out a tightness test (see Technical
tery or replace with
Manual)
fully charged bat-
tery.
Carry out a functional test (see "Check
the flow rate and warning devices",
page 24)
Replace the O-ring at the plug-in or bay-
onet-type hose connector (see "Replac-
Change particle or
ing the O-ring at plug-in or bayonet-type
combination filter.
connectors of breathing hoses",
page 24)
1) For gas-tight packed devices, otherwise every 6 months
2) Not required by Dräger. Observe national guidelines where applica-
Re-check the
device function
and prepare for
use.
Cause
General system
error
Cause
General error or
defect
Maintenance
Maintenance intervals
2)
ble.
Instructions for use
Remedy
Device must be
checked by Dräger
Service.
Remedy
Re-insert the
rechargeable bat-
tery in the battery
charger. If the error
occurs repeatedly
have Dräger Ser-
vice check the bat-
tery charger and
rechargeable bat-
tery.
1)
X
1)
X
1)
X
1)
X
X
|