11.0 Notes on EMC
■ Medical electrical equipment is subject to special precautions with regard to EMC and must be installed acc.
to following EMC notes.
■ Portable and mobile HF communication facilities can influence medical electrical equipment.
■ The use of other accessories, other converters and cables than stated may lead to an increased emission
or a reduced interference immunity of the equipment or system.
11.1 Guidelines and Manufacturer´s Declaration - Emissions
The ENT-camera Varioair 3 is intended for use in the electromagnetic environment specified below. The customer or
user of the Varioair 3 should ensure that it is used in such an environment.
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonics IEC 61000-3-2
Flicker
IEC 61000-3-3
The device may not be used directly next to other devices or piled up with other devices.
If operation next to or piled with other devices is necessary, please watch the device to check its intended
operation in this arrangement.
11.2 Guidelines and Manufaturer´s Declaration - Immunity
The ENT-camera Varioair 3 is intended for use in the electromagnetic environment specified below. The customer or
user of the Varioair 3 should ensure that it is used in such an environment.
Immunity Test
ESD
IEC 61000-4-2
EFT
IEC 61000-4-4
Surges
IEC 61000-4-5
Power Frequency 50/60
Hz
Magnetic field
IEC 61000-4-8
48
Compliance
Group 1
Class B
Class A
Match
IEC 60601-
Test Level
Electromagnetic Environment - Guidance
The ATMOS Varioair 3 uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference
in nearby electronic equipment.
The ATMOS Varioair 3 is suitable for use in all es-
tablishments, including domestic, and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Compliance Level
± 6 kV Contact
± 8 kV Air
± 2 kV Mains
± 1 kV I/Os
± 1 kV Differential
± 2 kV Common
3 A/m
Electromagnetic Environ-
ment - Guidance
Floors should be wood, concrete,
or ceramis tile. If floors are synthe-
tic, the relative humidity should be
at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Power frequency magnetic fields
should be that of a typical com-
mercial or hospital environment.
UK