Trouble Shooting - Atmos Varioair 3 Mode D'emploi

Masquer les pouces Voir aussi pour Varioair 3:
Table des Matières

Publicité

7.0
Trouble-shooting
Error in temperature display
"F0"
Not used
maxium allowed temperature
"F1"
exceeded (>51°C); if the tempera-
ture exceeds 51°C for more than 2
seconds, the pump is switched off
automatically
"F2"
-5V is missing (supply voltage on the
control board)
"F3"
Break of the safety NTC
"F4"
Not used
"F5"
Break of the regulating NTC
"F6"
Not used
"F7"
Temperature too high (>48°C)
"F8"
Short-circuit of the regulating NTC
"F9"
Not used
If errors cannot be corrected with the assistance of the trouble-shooting list, please inform the ATMOS service staff.
Do not attempt to repair the unit yourself!
36
UK
Switch off device, cooling down for approx. 1 minute
is required.
Check, whether temperature setting is too high.
If necessary, adjust desired temperature to a value of <51°C
by means of the coding switches.
Inform the service staff.
Bent hose tip within the ear/auditory canal.
Inform the service staff.
Check proper connection of the handle
Replace the handle
Inform the service staff.
Check proper connection of the handle
Replace the handle
Inform the service staff.
Check, whether temperature setting is too high.
If necessary, adjust desired temperature to a value of <48°C
by means of the coding switches.
Inform the service staff.
Bent hose tip within the ear/auditory canal.
Replace the handle
Have temperature feeler of the regulating NTC checked by
the service staff.
F
Page 37

Publicité

Table des Matières
loading

Table des Matières