7.0
Trouble-shooting
Error in temperature display
"F0"
Not used
maxium allowed temperature
"F1"
exceeded (>51°C); if the tempera-
ture exceeds 51°C for more than 2
seconds, the pump is switched off
automatically
"F2"
-5V is missing (supply voltage on the
control board)
"F3"
Break of the safety NTC
"F4"
Not used
"F5"
Break of the regulating NTC
"F6"
Not used
"F7"
Temperature too high (>48°C)
"F8"
Short-circuit of the regulating NTC
"F9"
Not used
If errors cannot be corrected with the assistance of the trouble-shooting list, please inform the ATMOS service staff.
Do not attempt to repair the unit yourself!
36
UK
•
Switch off device, cooling down for approx. 1 minute
is required.
•
Check, whether temperature setting is too high.
If necessary, adjust desired temperature to a value of <51°C
by means of the coding switches.
•
Inform the service staff.
•
Bent hose tip within the ear/auditory canal.
•
Inform the service staff.
•
Check proper connection of the handle
•
Replace the handle
•
Inform the service staff.
•
Check proper connection of the handle
•
Replace the handle
•
Inform the service staff.
•
Check, whether temperature setting is too high.
If necessary, adjust desired temperature to a value of <48°C
by means of the coding switches.
•
Inform the service staff.
•
Bent hose tip within the ear/auditory canal.
•
Replace the handle
•
Have temperature feeler of the regulating NTC checked by
the service staff.