6.0 Maintenance and service

Change of filter (see section 6.4) should be carried out immediately by the user when required.
Saftey-technical inspection
6.1

Functional test

Test
Switch unit several times on and off.
Correct function of mains pilot lamp.
Automatic activation of displays test.
Testing the displays.
All optical and audible indicators issue their signal during the first second after switching on (in particular seg-
mental test "8 8 ").
Set all temperature setting switches on 40°C and test consecutively by touching the corresponding keys whether
in all ranges 40°C is reached in ACTUAL temperature indication.
• Check air temperature.
• Carry out the same test with temperatures 37°C and 28°C.
The particular combination of temperatures 40°C, 37°C and 28°C, as well as the test in all ranges ensures
a comprehensive test of all contacts of the pre-setting switches, in ON- as well as in OFF-position, and also
testing of all components involved in temperature control of the DESIRED temperature control system.
Therefore, carrying out a test in one range only, or at only one temperature is not permissible.
Set temperature at 49°C and close the air exhaust opening on the handle.
The electronic excess-temperature protection must respond after approx. 100-120 sec. (in the temperature
display error code "F7" appears).
Measure temperature of air at standard settings: 30° and 44°C.
The value may not deviate more than ± 0,5° ± 1/2 digit, i.e. 1°C in total. Typically, however, values are
better than ± 0,5°C.
This measurement is critical !
Measuring probes or thermometers which are too large may cause faulty measurements because of substantial
heat removal, inertia, and because their temperature sensitivity is not localised on a point but spread over a
larger surface. Use measuring instruments with an immersion probe, having a minimum precision of 0.3°C and a
resolution of 0.1°C.
After setting the temperature, wait at least 1 minute before measuring to allow temperature stabilization of the
double-barrelled hose and handle.
After setting the temperature 1 minute at least should be waited in order that the probe attains correct
temperature.