5.2 Factory-set Alarm Thresholds and Setpoints
Sensor
LOW
HIGH
alarm
alarm
LEL
10% LEL
20% LEL
19.5%
23.0%
O 2
CO
25 ppm
100 ppm
10 ppm
15 ppm
H 2 S
This instrument is not approved for use in atmospheres containing >21 % oxygen.
5.3 Certifications
See instrument label for the approvals that apply to your specific unit.
USA and Canada (Non-mining)
See instrument label for approvals that apply to your specific unit
USA (Mining)
MSHA 30 CFR Part 22, Methane Detector.
Approved for 30 CFR Part 75 Determination testing for methane and oxygen deficiency
Europe Union
The product ALTAIR 4X complies with the following directives, standards or
standardized documents:
Directive 94/9/EC (ATEX): II 1G Ex ia IIC T4, Ga IP67
CE 0080
Directive 2004/108/EEC (EMC): EN50270:2007 Type 2, EN61000-6-3:2007
Australia/New Zealand Test Safe Australia
Ex ia sa I IP67 (Zone 0)
Ex ia sa IIC T4 IP67 (Zone 0)
Ta = -40 °C to +60 °C
IECEx TestSafe Australia
Ex ia I IP67 (Zone 0)
Ex ia mb d IIC T4 IP67 (Zone 1) - When Combustible XCell Sensor is installed
Ex ia IIC T4 IP67 (Zone 0) - When Combustible XCell Sensor is not installed
Ta = -40 °C to +60 °C
5-2
STEL
TWA
--
--
--
--
100
25
15
10
(Zone 0 with no combustible sensor installed)
II 2G Ex d ia mb IIC T4 Gb, IP67
(Zone 1 with combustible sensor installed)
I M1 Ex ia I Ma (Zone 0)
Ta = -40 °C to +60 °C
EN60079-29-1 ATEX Flammable Gas Performance
Certification for Groups I and II
EN50104 ATEX Oxygen Performance Certification
Min
Max
Auto-cal
alarm
alarm
values
setpoint
setpoint
5
60
58%
5
29
15.0%
15
1700
60 ppm
5
175
20 ppm