4.1 Troubleshooting
PROBLEM
DESCRIPTION
ERROR TEMP
Temperature is below
-40°F (-40°C) or
above 167°F (75°C)
ERROR EE
EEPROM Memory error
ERROR PRG
Flash Memory error
ERROR RAM
RAM Memory error
ERROR UNK
Unknown error
Battery warning repeats
LOW
every 15 seconds
BATTERY
Battery is completely
BATTERY
discharged
ALARM
ERROR
Charge error
CHARGE
SENSOR
Missing Sensor
ERROR
Instrument does
Low battery
not turn ON
Sensor warning
Sensor alarm
&
4-2
REACTION
Return to normal temperature range
and recalibrate.
Contact MSA
Contact MSA
Contact MSA
Contact MSA
Contact MSA
Remove from service as soon as
possible and recharge battery
Instrument is no longer sensing
gas. Remove from service
and recharge battery
Instrument must be between
10 and 35°C to charge.
Contact MSA if problem persists
Verify sensor is properly installed
and recalibrate.
Contact MSA if problem persists
Charge instrument
Sensor is near its end of life
Sensor has reached end of life
and cannot be calibrated.
Replace sensor and recalibrate