4.3 Cleaning the Instrument
Clean the exterior of the instrument regularly using only a damp cloth.
Do not use cleaning agents as many contain silicones which damage
the combustible sensor.
4.4 Storage
When not in use, store the instrument in a safe, dry place between
65°F and 86°F (18°C and 30°C). After storage, always recheck
instrument calibration before use.
4.5 Shipment
Pack the instrument in its original shipping container with suitable
padding. If the original container is unavailable, an equivalent container
may be substituted.