1.2. Safety and Precautionary Measures
Carefully review the following safety limitations and
precautions before placing this instrument in service.
Incorrect use can cause serious personal injury or death.
•
Check function (see section 3.6) each day before use. MSA
recommends carrying out a routine inspection prior to each day's
use.
•
Perform a Bump Test (see section 3.8) before each day's use to
verify proper instrument operation. The instrument must pass the
bump test. If it fails the test, perform a calibration (see section 3.9)
before using the instrument.
•
Perform a bump test more frequently if the unit is subjected to
physical shock or high levels of contaminants. Also, perform a
bump test more frequently if the tested atmosphere contains the
following materials, which may desensitize the combustible gas
sensor and reduce its readings:
•
Organic silicones
•
Silicates
•
Lead-containing compounds
•
Hydrogen sulfide exposures over 200 ppm or exposures
over 50 ppm for one minute.
•
The minimum concentration of a combustible gas in air that can
ignite is defined as the Lower Explosive Limit (LEL). A combustible
gas reading of "XXX" indicates the atmosphere is above 100 %
LEL or 5.00 %vol CH 4 , and an explosion hazard exists. Move
away from hazardous area immediately.
•
Do not use the ALTAIR 4X Multigas Detector to test for combustible
or toxic gases in the following atmospheres as this may result in
erroneous readings:
• Oxygen-deficient or oxygen-rich atmospheres
• Reducing atmospheres
• Furnace stacks
• Inert environments
• Atmospheres containing combustible airborne mists/dusts.
" WARNING
1-3