To use Delta (sampling) or Mean (sampling) mode,
the channel measurement mode must be configu-
red, then the sampling function selected for the
desired external contact, for example pedal 1.
In this configuration, every time pedal 1 is pres-
sed the channel value is temporarily recorded (n
indicates the number of values recorded) and the
delta or mean is calculated as a function of the
recorded values. This allows a delta to be pro-
duced at several points using a single instrument.
Channel calibration at 2 points
It is possible to calibrate the channel at two reference
points. A multiplication factor is then calculated so
allowing the measurement of diameters on V or the
calibration of a channel depending on the reference
block. This multiplication factor can then be activated
or deactivated. This calibration only affects the chan-
nel and not the instrument allocated to the channel.
System setting
Digital outputs
It is possible to configure the digital outputs as a function of the tolerance of a
specific channel, or generally.
Function: allows configuration of which condition activates the output:
NO GO channel: when the channel value is outside tolerances
•
GO channel: when the channel value is within tolerances
•
Global Tol <: when the general measurement is less than the defined tolerances
•
This menu enables access to the unit's
global configuration :
•
Display configuration
•
Digital output configuration
•
Configuration recording
•
Recorded measures visualisation
•
Modify printing options
14