Télécharger Imprimer la page

Example To Connect An External Contact On A Switch 1 Input; Rs485/D30X And Rs485/Mb-X Connections - sylvac D300S V2 Manuel D'utilisation

Masquer les pouces Voir aussi pour D300S V2:

Publicité

Les langues disponibles

Les langues disponibles

Example to connect a relay on the n°1 digital output
(internal power supply)
Note: a protective diode must be added in parallel of the relay if it isn't integrated.
Example to connect an external contact on the Switch 1 input

RS485/D30X and RS485/MB-X connections

These two inputs allow an increase in the number of probes connected to the
D300S.
The RS485/D30X input
This input is exclusively used for the
D302 and D304 modules.
The capacitive probes (P5, P10,
P25, P50) can be connected to these
modules.
Remark : the configuration jumpers must be placed on JP1 and JP3 so that the
module is correctly detected by the D300S.
The RS485/MB-X input (only available on certain versions)
This input is used to connect the MB-8i, MB-2C, MB-4C or even MB-2S type
modules. Inductive capacitive and incremental probes can be connected to
these modules.
NO
NC
COM
9