EXTxDELAYy:z
(x = 1 to 4)
(y = 1 to 3)
(z = 100 to 99999999 )
KEY0
KEY1
STO?x
(x = name listed in the file menu of the va-
lues recorded)
VIEWxy
(x = 1 to 5 and y = file name)
#xx#+...
(xx : 1 to 64)
#xx#RES1
#xx#RES2
#xx#RES3
#xx#RES4
#xx#BAR
#xx#NUM
#xx#MAX
#xx#MIN
#xx#DEL
#xx#NOR
#xx#MM
#xx#IN
#xx#PRE?
Ex: « EXT3FCT1:1 » configures pedal 1 with the Preset func-
tion
Introduction of a delay in [ms] between the execution of the
different functions of the external contact.
X = 1: switch 1
Y=1: delay 1
X = 2: switch 2
Y=2: delay 2
X = 3: Foot pe-
Y=3: delay 3
dal 1
X = 4: Foot Pe-
dal 2
Ex : « EXT3DELAY1 :1000 » introduction of a 1000ms delay
between the execution of function 1 and 2 of pedal 1.
Activates the keyboard and tactile screen
Locks the keyboard and tactile screen
Returns recorded values of the requested channel
Only works in static mode
Select the measure mode or the static mode
X = 1: measure mode
X = 2: displays the histogram (statistic mode)
X = 3: displays the chart on average (statistic mode)
X = 4: displays the chart on the range (static mode)
X = 5: displays the statistics table (statistics mode)
Ex : « VIEW2channel1 » displays the channel1 histogram.
All the following retro-commands are preceded by « #xx# ».
This means you must specify the channel on which the com-
mand is carried out.
If the channel number is not specified, the command is exe-
cuted on all active channels.
Changes the resolution: 0.0001
Changes the resolution: 0.001
Changes the resolution: 001
Changes the resolution: 0.1
Selects the bargraph display mode
Selects the digital display mode
Selects the maximum measure mode
Selects the minimum measure mode
Selects the delta measure mode (max-min)
Selects the normal measure mode
Sets the measure unit in millimetres
Sets the measure unit in inches
Returns the Preset value
25
Z=9: Hold OFF
Z=10: Hold ON/
OFF
Z = time in ms
(min value = 100 ms)
E