«Measurement screen» window
Item
Read
command
Zero
Clear
Preset
Print
Part
Switch off
Display calibration / Pairing with a probe
The D70 for inductive probes is delivered calibrated with the desired probe type.
For best results it is possible to recalibrate the display with your reference sensor to have ho-
mogeneity in your set of displays or to pair a sensor and display assembly.
Attention: After this procedure, the original calibration of your D70 will be irretrievably lost and
replaced by the values from your operation. This function should therefore be used with care.
For this, follow the procedure below:
1. Switch off the D70
2. Switch on the D70
3. When the start screen appers, press Sylvac Logo
4. A desktop with five icons appears, press
«Calibration».
5. The following screen appears
The probe position is displayed:
-
Position the value at zero
-
Place a 1mm gauge block under the probe
(or move the probe 1mm using a
precision movement system).
-
Press OK
-
The D70 is now recalibrated
Press the Home icon to return to the standard
D70 menu.
Write
Comment
command
ZERO
Relative zero
CLR
Dynamic measurement
PRESET
Calibration
?
G1
Part 1
G2
Part 2
OFF
Switches the equipement off
20