15.1.2. Guidance and Manufacturer's Declaration: Electromagnetic Immunity
Zen-O lite™ is intended for use in the electromagnetic environment specifi ed below. The
customer or the user of the concentrator should assure that it is used in such an environ-
ment.
IMMUNITY
IEC 60601 TEST
TEST
Electrostatic
± 2, 4, 6, 8kV
discharge
contact
(ESD)
± 2, 4, 8, 10, 15kV
IEC 61000-4-2
air
Electric fast
± 2kV for power
transient/burst
supply lines
IEC 61000-4-4
± 1kV for input/
output lines
Surge
± 1kV diff erential
IEC 61000-4-5
mode
± 2kV common
mode
Voltage dips,
< 5% UT
short inter-
(>95% dip in UT)
ruptions,
for 0.5 cycle
and voltage
40% UT
variations
(60% dip in UT)
on power
for 5 cycles
supply
70% UT
input lines
(30% dip in UT)
IEC 61000-4-11
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
Power
30 A/m
frequency
(50/60 Hz) ma-
gnetic fi eld
IEC 61000-4-8
COMPLIANCE
LEVEL
LEVEL
± 2, 4, 6, 8kV
contact
± 2, 4, 8, 10, 15kV
air
± 2kV for power
supply lines
± 1kV for input/
output lines
± 1kV diff erential
mode
± 2kV common
mode
< 5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
30 A/m
39/362
ELECTROMAGNETIC
ENVIRONMENT/GUIDANCE
Floors should be wood, concrete, or
ceramic tile.
If fl oors are covered with synthetic
material, the relative humidity should
be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should
be that of a typical commercial or
hospital environment. If the user
of the Zen-O  lite™ Portable Oxygen
Concentrator required continued
operation during power main inter-
ruptions, it is recommended that
the concentrator be powered from
an uninterruptible power supply or
battery.
Power frequency magnetic fi elds
should be at levels characteristic
of a typical commercial or hospital
environment.
EN